How to Calibrate a Pipette (Ultra quality)

This calibration log documents how to calibrate a pipette using the gravimetric method, in which distilled water is aspirated and dispensed onto a precision balance to verify accuracy and precision. Calibrating a pipette correctly guarantees maximum reliability and confidence in your laboratory results, and depends on three key factors: the equipment used, the procedures followed, and the environment in which calibration is performed. Use this log to record instrument details, the calibration procedure, reference standards, measured results, maintenance activity, and final approval for release to use.

Lab· 4 steps· 11 screenshots· 1569 words· Source video 3:51

Video: Mastering Pipette Calibration: Ensuring Accuracy and Precision by Sartorius (2014). All credit for the demonstration goes to the creator; watch the original on YouTube. The written guide below was generated from this video by Docsie. Creator? Request a change or removal.

This calibration log documents how to calibrate a pipette using the gravimetric method, in which distilled water is aspirated and dispensed onto a precision balance to verify accuracy and precision. Calibrating a pipette correctly guarantees maximum reliability and confidence in your laboratory results, and depends on three key factors: the equipment used, the procedures followed, and the environment in which calibration is performed. Use this log to record instrument details, the calibration procedure, reference standards, measured results, maintenance activity, and final approval for release to use.

A laboratory technician in a white lab coat and blue gloves pre-wetting a pipette tip at a calibration station. The workspace includes a balance, pipette stand, disposable tip boxes, and a laptop.
A laboratory technician in a white lab coat and blue gloves pre-wetting a pipette tip at a calibration station. The workspace includes a balance, pipette stand, disposable tip boxes, and a laptop.
Close-up of a laboratory balance display showing a measurement of 0.04965 g. The display includes additional information such as maximum capacity and calibration status.
Close-up of a laboratory balance display showing a measurement of 0.04965 g. The display includes additional information such as maximum capacity and calibration status.

Instrument details

Field Details
Instrument ID Not specified in this session; assign the pipette's unique asset tag before logging results.
Model Single-channel pipette
Serial Number Not specified in this session; record the serial number from the pipette body before starting.
Location Laboratory calibration station
Owner/Custodian Laboratory technician assigned to operate and maintain the pipette
Intended Use Precise aspiration and dispensing of liquid volumes, verified using distilled water on a gravimetric balance

The calibration station is set up on a stable, vibration-free surface, with distilled water and disposable tips kept within easy reach. The calibration software runs on a nearby laptop or PC for data entry and analysis, and pipettes are organized on a stand to keep them accessible and free of contamination.

A laboratory bench setup with a balance, distilled water in a beaker, disposable pipette tips in boxes, a pipette stand with multiple pipettes, a waste container, and a laptop running calibration software. Each item is labeled: Balance, Distilled Water, Disposable Tips, Calibration Software.
A laboratory bench setup with a balance, distilled water in a beaker, disposable pipette tips in boxes, a pipette stand with multiple pipettes, a waste container, and a laptop running calibration software. Each item is labeled: Balance, Distilled Water, Disposable Tips, Calibration Software.

Calibration schedule

Field Details
Frequency Not specified in this session; follow your laboratory's SOP or the manufacturer's recommended interval.
Due Date Not specified in this session; record the next due date after completing this calibration.
Procedure Gravimetric calibration: pre-wet the tip, aspirate vertically, tare the balance, dispense at a 45-degree angle, record the mass, and repeat for 3–5 replicates
Responsible Role Laboratory technician performing the calibration, with a second technician available to review documentation
A laboratory technician in a white lab coat and blue gloves pipetting at a calibration station, maintaining a steady rhythm. The workspace is organized and clean.
A laboratory technician in a white lab coat and blue gloves pipetting at a calibration station, maintaining a steady rhythm. The workspace is organized and clean.
Close-up of gloved hands inspecting the body and tip of a pipette for leakage or mechanical faults. The technician is using a calibration tool.
Close-up of gloved hands inspecting the body and tip of a pipette for leakage or mechanical faults. The technician is using a calibration tool.

Before you begin, power on the balance and confirm it is tared (zeroed), and open the calibration software to prepare it for data entry. Wear appropriate personal protective equipment, such as lab gloves and a lab coat, and pre-wet the pipette tips to improve the reliability of your results.

A gloved hand holding a pipette, demonstrating proper handling technique in a laboratory environment. The workspace includes pipettes, a balance, disposable tips, and a laptop with calibration software.
A gloved hand holding a pipette, demonstrating proper handling technique in a laboratory environment. The workspace includes pipettes, a balance, disposable tips, and a laptop with calibration software.

Standards used

Field Details
Reference Standard Distilled water, used as the calibration liquid for gravimetric verification
Lot Number Not specified in this session; record the batch or lot of distilled water used.
Certificate An accredited calibration certificate must be generated on completion, including the uncertainty of measurement for the instrument
Expiration Date Not specified in this session; verify certificate validity per your accreditation body's requirements.
Traceability The certificate documents the results, the equipment used, and the personnel involved, providing full traceability of the calibration
Overhead view of a gloved technician holding a pipette at a 45-degree angle, dispensing liquid into a weighing vessel on a balance. Pipette tip boxes and calibration equipment are visible.
Overhead view of a gloved technician holding a pipette at a 45-degree angle, dispensing liquid into a weighing vessel on a balance. Pipette tip boxes and calibration equipment are visible.

Review the certificate details after calibration to confirm that all required information, including the measurement uncertainty, is present before accepting the pipette back into service.

Calibration results

Steps to calibrate a pipette using the gravimetric method

  1. Pre-wet the pipette tip. Aspirate and dispense distilled water three to five times before starting calibration. This eliminates evaporation of water inside the pipette and improves the accuracy of your results.

  2. Aspirate the water correctly. Hold the pipette vertically, depress the plunger to the first stop, immerse the tip in the water, and slowly release the plunger to draw up the sample. Avoid tilting the pipette during aspiration.

  3. Confirm the balance is tared. Before dispensing, verify that the balance display reads zero. Taring is often performed automatically by the calibration software, but you should still check it.

  4. Dispense the sample onto the balance. Hold the pipette at a 45-degree angle, position the tip against the wall of the receiving vessel, and press the plunger to the second stop to expel the entire sample.

  5. Record the measurement. Read the stable balance display and log the measured mass, for example 0.04831 g.

  6. Repeat for statistical reliability. Repeat aspiration and dispensing for the required number of replicates, typically 3–5 times. The smaller the sample volume, the more important it is to expel the entire sample and avoid residual droplets.

  7. Maintain consistent technique. Keep an even rhythm and timing for each pipetting cycle to reduce variability and improve the reliability of your results.

  8. Document the results. Enter all results into the calibration software or your laboratory notebook, and generate a calibration certificate recording the results, equipment, and personnel involved.

  9. Verify and finalize. Double-check that all measurements and documentation are complete and accurate, then save or print the calibration certificate for your records.

Technique and environmental checks during calibration

  1. Attach tips without contamination. Press the pipette straight down onto the tip without twisting. Use tip racks and avoid touching tips with your hands to maintain sterility.

  2. Control the calibration environment. Keep the calibration area draft-free and maintain constant temperature and humidity, since environmental changes can affect accuracy. Use environmental monitoring equipment where available.

  3. Apply consistent handling for every replicate. Hold the pipette vertically when aspirating and at a 45-degree angle when dispensing into a vessel on the balance, following the same procedure each time.

  4. Follow the best-practice sequence. Pre-wet the tip, aspirate vertically, dispense at a 45-degree angle, and repeat each step consistently for every calibration cycle.

  5. Dispense with care. Maintain the 45-degree angle and position the pipette tip correctly inside the vessel to avoid splashing or loss of sample.

  6. Avoid contact with the vessel walls. Dispense carefully without touching the sides of the vessel with the pipette tip, to prevent contamination or measurement errors.

Close-up of a pipette being pressed vertically into a yellow pipette tip rack, demonstrating correct tip attachment. Orange tip racks are also visible.
Close-up of a pipette being pressed vertically into a yellow pipette tip rack, demonstrating correct tip attachment. Orange tip racks are also visible.
Close-up of a laboratory technician in a white lab coat and glasses, focused on the calibration process.
Close-up of a laboratory technician in a white lab coat and glasses, focused on the calibration process.

Post-adjustment verification results

  1. Verify the set volume. Check the pipette's display to confirm the desired volume is set, for example 50.0 µL, before dispensing for verification.

  2. Dispense and assess precision. Dispense the set volume into the calibration vessel on the balance and observe the reading. Repeat dispensing multiple times to evaluate the distribution of dispensed volumes.

  3. Record and compare results. Log the measured mass for each cycle, for example 0.04965 g, and compare results to expected values to assess accuracy and precision.

  4. Confirm final reliability. Review the final balance reading, for example 0.04999 g, to confirm consistent and reliable performance before releasing the pipette for use.

Result summary table

Replicate Set Volume Measured Mass Acceptance Limit Pass/Fail Analyst Initials
Initial run Not specified in this session 0.04831 g Not specified in this session; define per your SOP Not specified in this session Not specified in this session
Post-adjustment 50.0 µL 0.04965 g Not specified in this session Not specified in this session Not specified in this session
Final verification 50.0 µL 0.04999 g Not specified in this session Not specified in this session Not specified in this session

Maintenance

Field Details
Preventive Maintenance Inspect the pipette body and tip for damage, wear, and proper sealing during each calibration
Repairs Not specified in this session
Parts Replaced Not specified beyond disposable tips used for each calibration cycle
Service Provider Not specified; adjustments follow the manufacturer's instructions using the designated calibration tool

Adjust the pipette volume as needed. Hold the pipette securely with one hand and use the calibration tool to turn the adjustment mechanism with the other. Turn clockwise to increase the dispensed volume and counterclockwise to decrease it, making adjustments gently and precisely.

Inspect the pipette for leakage or mechanical faults. Examine the pipette body and tip for visible damage or wear, check for proper sealing and smooth plunger operation, and use the calibration tool to inspect internal components if needed.

A technician in a white lab coat and blue gloves examining a pipette, preparing to check for faults. Another technician observes.
A technician in a white lab coat and blue gloves examining a pipette, preparing to check for faults. Another technician observes.

Out-of-tolerance actions

Impact assessment, affected data review, and quarantine. Not specified in this session; record the assessment of data generated since the last valid calibration, note any affected results, and quarantine the pipette from use until the issue is resolved.

If results indicate inaccuracy, use special calibration tools to adjust the pipette, following the manufacturer's instructions, then recalibrate and verify accuracy after the adjustment.

If accuracy cannot be corrected by adjustment, attempt further adjustments if necessary; if the pipette still cannot achieve accurate results, proceed to check for mechanical issues.

Reiterate the calibration factors. Confirm that every step was performed with the correct equipment, following standardized procedures, and in a controlled environment, since the technician's technique and consistency are critical for reliable results.

Approval

Field Details
Reviewer A second laboratory technician reviews the calibration results and documentation
Date Not specified in this session; record the date of final review
Release-to-Use Decision Confirm the pipette meets laboratory standards based on the final balance readings before returning it to service

Collaborate and review documentation. Work with a colleague to review the calibration results and documentation, and discuss any uncertainties or questions regarding the process or certificate, while maintaining a professional, organized laboratory environment.

A gloved hand holding a single-channel pipette, removing it from a pipette stand in a clean laboratory environment. Other pipettes are visible in the background.
A gloved hand holding a single-channel pipette, removing it from a pipette stand in a clean laboratory environment. Other pipettes are visible in the background.

Complete the calibration process. Save all data and update the calibration certificate, then return the pipette and calibration tools to their designated storage locations, maintaining a clean and organized laboratory environment.

What this guide cost to generate

Docsie billed 8,000 credits (about $5.60) to analyze this 4-minute video at ultra quality. The rewrite, template fill and Word/PDF exports were included. The same video at each quality tier:

QualityFrames sampledCreditsApprox. cost
Draftevery 16-30 s1,000$0.70
Standardevery 8-15 s2,000$1.40
Detailedevery 4-7 s4,000$2.80
Ultra (this guide)every 1-3 s8,000$5.60

Credits priced at $0.70 per 1,000; plans include a monthly allowance. Enterprise customers on on-premise or bring-your-own-model deployments run this on their own inference and pay no per-video credits.

Same video, other quality tiers

We ran this exact video through Docsie again at different frame-sampling rates. Compare the guides side by side:

QualityFrame everyScreenshotsCreditsCost
Ultra (this guide)1 s118,000$5.60
Standard8-15 s92,000$1.40
Download this filled Instrument Calibration Log: PDFWord (.docx) Blank template

Generated by Docsie Video-to-Docs on 2026-09-10 from a 3-minute video. It was generated straight into Docsie's Instrument Calibration Log template, so every section of that template is filled from the video. Screenshots are frames from the source video and belong to their creator, Sartorius, whose original is embedded above. If you own this video and want the guide removed or credited differently, contact us and we will act within one business day.

Video to Documentation

Have your own lab process videos?

Docsie turned a 3-minute YouTube video into the guide above, screenshots included, with no manual writing. Upload your Zoom, Teams, Loom or screen recordings and get the same for your team.

Steps + screenshots extracted automatically
Export to Markdown, DOCX, PDF or publish to a portal